Product Name:
Rat Zonula occludens protein 1 (ZO-1) ELISA Kit
Cat.#:
JL21101
Regulatory Status:
Alternative:
ZO1; Zona Occludens 1; Zona occludens protein 1; Zonula occludens protein 1
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.156-10ng/mL
Sensitivity:
0.074ng/mL
Specificity:
This kit recognizes Rat ZO-1 in samples. No significant cross-reactivity or interference between Rat ZO-1 and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:

Research Area:
Signal Transduction
Product Datasheets
| ZO-1
Tight junction protein ZO-1 also known as Zonula Occludens-1 (ZO-1), is a 220-kD peripheral membrane protein that is encoded by the TJP1 gene. It belongs to the family of zonula occludens proteins (ZO-1, ZO-2, and ZO-3), which are tight junction-associated proteins and of which, ZO-1 is the first to be cloned. It was first isolated in 1986 by Stevenson and Goodenough using a monoclonal antibody raised in rodent liver to recognise a 225-kD polypeptide in whole liver homogenates and in tight junction-enriched membrane fractions. It has a role as a scaffold protein which cross-links and anchors Tight Junction (TJ) strand proteins, which are fibril-like structures within the lipid bilayer, to the actin cytoskeleton.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.Concentration (ng/mL) | 10 | 5 | 2.5 | 1.25 | 0.625 | 0.312 | 0.156 | 0 |
OD | 2.44 | 1.87 | 1.29 | 0.83 | 0.58 | 0.38 | 0.27 | 0.08 |
Corrected OD | 2.36 | 1.79 | 1.21 | 0.75 | 0.5 | 0.3 | 0.19 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Rat ZO-1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.Sample Type | Range(%) | Average Recovery(%) |
Serum(n=8) | 84-101 | 96 |
Plasma(n=8) | 92-105 | 102 |
Cell Culture Supernates(n=8) | 96-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Rat ZO-1 and diluted with Universal Diluent to produce samples with values within the range of the assay.Serum(n=4) | Plasma(n=4) | Cell Culture Supernates(n=4) | ||
1 : 2 | Range (%) | 84-95 | 88-96 | 90-110 |
Average (%) | 91 | 93 | 96 | |
1 : 4 | Range (%) | 89-103 | 87-108 | 105-115 |
Average (%) | 94 | 98 | 108 |






Size
Quantity
Price ($)
48T


360 / 360
96T


460 / 460