Product Name:
Rat Angiotensin II Receptor 1 (AT1R) ELISA Kit
Cat.#:
JL21364
Regulatory Status:
Alternative:
AG2S; AG2-S; AGTR1A; AGTR1B; AT1; AT1B; AT2R1; AT2-R1; AT2R1A; AT2R1B; HAT1R; AT1 Receptor; Angiotensin Receptor 1B
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.15-10ng/mL
Sensitivity:
0.05ng/mL
Specificity:
This kit recognizes Rat AT1R in samples. No significant cross-reactivity or interference between Rat AT1R and analogues was observed.
Sample Type:
Tissue homogenates, cell lysates and other biological fluids
Assay Procedure:

Research Area:
Signal Transduction
Product Datasheets
| AT1R
Angiotensin II receptor type 1 (AT1) is a Gq/11-coupled G protein-coupled receptor (GPCR) and the best characterized angiotensin receptor. It is encoded by the AGTR1 gene. AT1 has vasopressor effects and regulates aldosterone secretion. It is an important effector controlling blood pressure and volume in the cardiovascular system. Angiotensin II receptor blockers are drugs indicated for hypertension, diabetic nephropathy and congestive heart failure.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.Concentration (ng/mL) | 10 | 5 | 2.5 | 1.25 | 0.62 | 0.31 | 0.15 | 0 |
OD | 2.15 | 1.63 | 1.13 | 0.71 | 0.5 | 0.33 | 0.23 | 0.07 |
Corrected OD | 2.08 | 1.56 | 1.06 | 0.64 | 0.43 | 0.26 | 0.16 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Rat AT1R spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.Sample Type | Range(%) | Average Recovery(%) |
Tissue Homogenates(n=8) | 92-105 | 101 |
Cell Lysates(n=8) | 96-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Rat AT1R and diluted with Universal Diluent to produce samples with values within the range of the assay.Tissue Homogenates(n=4) | Cell Lysates(n=4) | ||
1 : 2 | Range (%) | 84-95 | 90-110 |
Average (%) | 91 | 96 | |
1 : 4 | Range (%) | 89-103 | 105-115 |
Average (%) | 94 | 108 |






Size
Quantity
Price ($)
48T


360 / 360
96T


460 / 460