Product Name:
Rat G ApolipoProtein B (GApo-B) ELISA Kit
Cat.#:
JL52911
Regulatory Status:
Alternative:
GApo-B
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
31.2-2000ng/mL
Sensitivity:
15.9ng/mL
Specificity:
This kit recognizes Rat GApo-B in samples. No significant cross-reactivity or interference between Rat GApo-B and analogues was observed.
Sample Type:
Serum;plasma;tissue homogenates;cell lysates;cell culture supernates and other biological fluids
Assay Procedure:
Research Area:
Signal Transduction
Product Datasheets
| GApo-B
Apolipoprotein B is the primary apolipoprotein of chylomicrons, VLDL, Lp(a), IDL, and LDL particles (LDL—commonly known as "bad cholesterol" when in reference to both heart disease and vascular disease in general), which is responsible for carrying fat molecules (lipids), including cholesterol, around the body to all cells within all tissues. While all the functional roles of ApoB within the LDL (and all larger) particles remain somewhat unclear, it is the primary organizing protein (of the entire complex shell enclosing/carrying fat molecules within) component of the particles and is absolutely required for the formation of these particles.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (ng/mL) | 2000 | 1000 | 500 | 250 | 125 | 62.5 | 31.2 | 0 |
| OD | 2.22 | 1.56 | 0.92 | 0.63 | 0.42 | 0.25 | 0.18 | 0.08 |
| Corrected OD | 2.14 | 1.48 | 0.84 | 0.55 | 0.34 | 0.17 | 0.1 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Rat GApo-B spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 84-101 | 96 |
| Plasma(n=8) | 92-105 | 102 |
| Cell Culture Supernates(n=8) | 96-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Rat GApo-B and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | Cell Culture Supernates(n=4) | ||
| 1 : 2 | Range (%) | 84-95 | 88-96 | 90-110 |
| Average (%) | 91 | 93 | 96 | |
| 1 : 4 | Range (%) | 89-103 | 87-108 | 105-115 |
| Average (%) | 94 | 98 | 108 |
Size
Quantity
Price ($)
48T
1
280 / 280
96T
1