Product Name:
Rat Low Density Lipoprotein (LDL) ELISA Kit
Cat.#:
JL13846
Regulatory Status:
Alternative:
Low Density Lipoprotein
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
7.8-500ng/mL
Sensitivity:
3.55ng/mL
Specificity:
This kit recognizes Rat LDL in samples. No significant cross-reactivity or interference between Rat LDL and analogues was observed.
Sample Type:
Serum, plasma and other biological fluids
Assay Procedure:

Research Area:
Metabolism
Product Datasheets
| LDL
Low-density lipoprotein (LDL) is one of the five major groups of lipoprotein that transport all fat molecules around the body in extracellular water.These groups, from least dense to most dense, are chylomicrons (aka ULDL by the overall density naming convention), very low-density lipoprotein (VLDL), intermediate-density lipoprotein (IDL), low-density lipoprotein (LDL) and high-density lipoprotein (HDL). LDL delivers fat molecules to cells. LDL has been associated with the progression of atherosclerosis.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.Concentration (ng/mL) | 500 | 250 | 125 | 62.5 | 31.25 | 15.62 | 7.8 | 0 |
OD | 2.48 | 1.65 | 1.07 | 0.71 | 0.45 | 0.28 | 0.19 | 0.09 |
Corrected OD | 2.39 | 1.56 | 0.98 | 0.62 | 0.36 | 0.19 | 0.1 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Rat LDL spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.Sample Type | Range(%) | Average Recovery(%) |
Serum(n=8) | 84-101 | 96 |
Plasma(n=8) | 92-105 | 102 |
| Linearity
Samples were spiked with high concentrations of Rat LDL and diluted with Universal Diluent to produce samples with values within the range of the assay.Serum(n=4) | Plasma(n=4) | ||
1 : 2 | Range (%) | 84-95 | 88-96 |
Average (%) | 91 | 93 | |
1 : 4 | Range (%) | 89-103 | 87-108 |
Average (%) | 94 | 98 |






Size
Quantity
Price ($)
48T


360 / 360
96T