400-0066-400

Rat B-Cell Leukemia/Lymphoma 2 (Bcl-2) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Rat B-Cell Leukemia/Lymphoma 2 (Bcl-2) ELISA Kit
Product Name:

Rat B-Cell Leukemia/Lymphoma 2 (Bcl-2) ELISA Kit

Cat.#:

JL10037

Regulatory Status:
Alternative:

B-Cell CLL/Lymphoma 2

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.78-50ng/mL
Sensitivity:
0.28ng/mL
Specificity:
This kit recognizes Rat Bcl-2 in samples. No significant cross-reactivity or interference between Rat Bcl-2 and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Immunology
  • Rat B-Cell Leukemia/Lymphoma 2 (Bcl-2) ELISA Kit
  • Rat B-Cell Leukemia/Lymphoma 2 (Bcl-2) ELISA Kit
  • Rat B-Cell Leukemia/Lymphoma 2 (Bcl-2) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| Bcl-2

B-Cell Leukemia (Bcl-2) is a founding member of the regulatory protein Bcl-2 family, which blocks programmed cell death (apoptosis). BCL-2 is localized to the outer membrane of mitochondria, where it plays an important role in promoting cellular survival and inhibiting the actions of pro-apoptotic proteins. The pro-apoptotic proteins in the BCL-2 family, including Bax and Bak, normally act on the mitochondrial membrane to promote permeabilization and release of cytochrome c and ROS, that are important signals in the apoptosis cascade.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 50 25 12.5 6.25 3.12 1.56 0.78 0
OD 2.56 1.45 0.8 0.48 0.33 0.23 0.2 0.08
Corrected OD 2.48 1.37 0.72 0.4 0.25 0.15 0.12 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Rat Bcl-2 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-101 95
Plasma(n=8) 92-105 102
Cell Culture Supernates(n=8) 96-108 105





 

| Linearity

Samples were spiked with high concentrations of Rat Bcl-2 and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Culture Supernates(n=4)
1 : 2 Range (%) 84-95 88-96 90-110
Average (%) 91 93 96
1 : 4 Range (%) 89-103 87-108 105-115
Average (%) 94 98 108







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
360 / 360
96T
1
460 / 460
Top