400-0066-400

Mouse Aryl Hydrocarbon Receptor (AhR) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Mouse Aryl Hydrocarbon Receptor (AhR) ELISA Kit
Product Name:

Mouse Aryl Hydrocarbon Receptor (AhR) ELISA Kit

Cat.#:

JL20328

Regulatory Status:
Alternative:

BHLHE76; Class E basic helix-loop-helix protein 76

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.15-10ng/mL
Sensitivity:
0.07ng/mL
Specificity:
This kit recognizes Mouse AhR in samples. No significant cross-reactivity or interference between Mouse AhR and analogues was observed.
Sample Type:
tissue homogenates, cell lysates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Signal Transduction
  • Mouse Aryl Hydrocarbon Receptor (AhR) ELISA Kit
  • Mouse Aryl Hydrocarbon Receptor (AhR) ELISA Kit
  • Mouse Aryl Hydrocarbon Receptor (AhR) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| AhR

The aryl hydrocarbon receptor (also known as AhR, AHR, ahr, ahR, AH receptor, or as the dioxin receptor) is a protein that is encoded by the AHR gene. The aryl hydrocarbon receptor is a transcription factor that regulates gene expression. It was originally thought to function primarily as a sensor of xenobiotic chemicals and also as the regulator of enzymes such as cytochrome P450s that metabolize these chemicals. The most notable of these xenobiotic chemicals are aromatic (aryl) hydrocarbons from which the receptor derives its name.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 10 5 2.5 1.25 0.62 0.31 0.15 0
OD 2.28 1.24 0.67 0.35 0.19 0.14 0.11 0.09
Corrected OD 2.19 1.15 0.58 0.26 0.1 0.05 0.02 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Mouse AhR spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Tissue Homogenates(n=8) 84-101 96
Cell Lysates(n=8) 92-105 102





 

| Linearity

Samples were spiked with high concentrations of Mouse AhR and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Tissue Homogenates(n=4) Cell Lysates(n=4)
1 : 2 Range (%) 84-96 88-96
Average (%) 91 93
1 : 4 Range (%) 89-103 87-108
Average (%) 94 98







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
360 / 360
96T
1
460 / 460
Top