Product Name:
Mouse Sclerostin (SOST) ELISA Kit
Cat.#:
JL24692
Regulatory Status:
Alternative:
VBCH; Sclerosteosis
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
6.25-400pg/mL
Sensitivity:
3.5pg/mL
Specificity:
This kit recognizes Mouse SOST in samples. No significant cross-reactivity or interference between Mouse SOST and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenate, cell lysate, cell culture supernatant and other biological fluids
Assay Procedure:
Research Area:
Metabolism
Product Datasheets
| SOST
Sclerostin is a protein that is encoded by the SOST gene.It is a secreted glycoprotein with a C-terminal cysteine knot-like (CTCK) domain and sequence similarity to the DAN family of bone morphogenetic protein (BMP) antagonists. Sclerostin is produced primarily by the osteocyte but is also expressed in other tissues,and has anti-anabolic effects on bone formation.Sclerostin has been identified as binding to LRP5/6 receptors and inhibiting the Wnt signaling pathway.The inhibition of the Wnt pathway leads to decreased bone formation.Although the underlying mechanisms are unclear, it is believed that the antagonism of BMP-induced bone formation by sclerostin is mediated by Wnt signaling, but not BMP signaling pathways.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (pg/mL) | 400 | 200 | 100 | 50 | 25 | 12.5 | 6.25 | 0 |
| OD | 2.22 | 1.56 | 0.92 | 0.63 | 0.42 | 0.25 | 0.18 | 0.08 |
| Corrected OD | 2.14 | 1.48 | 0.84 | 0.55 | 0.34 | 0.17 | 0.1 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Mouse SOST spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 84-101 | 96 |
| Plasma(n=8) | 92-105 | 102 |
| Cell Culture Supernates(n=8) | 96-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Mouse SOST and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | Cell Culture Supernates(n=4) | ||
| 1 : 2 | Range (%) | 84-95 | 88-96 | 90-110 |
| Average (%) | 91 | 93 | 96 | |
| 1 : 4 | Range (%) | 89-103 | 87-108 | 105-115 |
| Average (%) | 94 | 98 | 108 |
Size
Quantity
Price ($)
48T
1
280 / 280
96T
1
460 / 460