Product Name:
Mouse Anti-tartrate Acid Phosphatase 5b (TRACP-5b) ELISA Kit
Cat.#:
JL11586
Regulatory Status:
Alternative:
TRACP-5b
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.078-5mIU/mL
Sensitivity:
0.029mIU/mL
Specificity:
This kit recognizes Mouse TRACP-5b in samples. No significant cross-reactivity or interference between Mouse TRACP-5b and analogues was observed.
Sample Type:
Serum;plasma;tissue homogenates and other biological fluids
Assay Procedure:
Research Area:
Immunology
Product Datasheets
| TRACP-5b
TRACP-5b is a subtype of tartrate-resistant acid phosphatase (TRAP) and belongs to the type 5 isoenzyme of the acid phosphatase family, with a molecular weight of approximately 30-40 kDa. It is specifically secreted by osteoclasts and serves as a critical biomarker reflecting bone resorption activity.Involved in osteopontin/bone sialoprotein dephosphorylation. Its expression seems to increase in certain pathological states such as Gaucher and Hodgkin diseases, the hairy cell, the B-cell, and the T-cell leukemias.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (mIU/mL) | 5 | 2.5 | 1.25 | 0.625 | 0.312 | 0.156 | 0.078 | 0 |
| OD | 2.08 | 1.58 | 1.06 | 0.7 | 0.43 | 0.3 | 0.23 | 0.08 |
| Corrected OD | 2 | 1.5 | 0.98 | 0.62 | 0.35 | 0.22 | 0.15 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Mouse TRACP-5b spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 84-101 | 96 |
| Plasma(n=8) | 92-105 | 102 |
| Tissue Homogenates(n=8) | 96-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Mouse TRACP-5b and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | Tissue Homogenates(n=4) | ||
| 1 : 2 | Range (%) | 84-95 | 88-96 | 90-111 |
| Average (%) | 91 | 93 | 96 | |
| 1 : 4 | Range (%) | 89-103 | 87-108 | 105-115 |
| Average (%) | 94 | 98 | 108 |
Size
Quantity
Price ($)
48T
1
280 / 280
96T
1
460 / 460