400-0066-400

Mouse Plasminogen Activator, Urokinase (uPA) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Mouse Plasminogen Activator, Urokinase (uPA) ELISA Kit
Product Name:

Mouse Plasminogen Activator, Urokinase (uPA) ELISA Kit

Cat.#:

JL10955

Regulatory Status:
Alternative:

PLAU; ATF; URK; UK; UP-A; Abbokinase; Urokinase-Type Plasminogen Activator;Plasminogen Activator, Urokinase

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.31-20ng/mL
Sensitivity:
0.12ng/mL
Specificity:
This kit recognizes Mouse uPA in samples. No significant cross-reactivity or interference between Mouse uPA and analogues was observed.
Sample Type:
Serum, plasma and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Cancer
  • Mouse Plasminogen Activator, Urokinase (uPA) ELISA Kit
  • Mouse Plasminogen Activator, Urokinase (uPA) ELISA Kit
  • Mouse Plasminogen Activator, Urokinase (uPA) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| uPA

Urokinase, also known as urokinase-type plasminogen activator (uPA), is a serine protease present in humans and other animals. The human urokinase protein was discovered, but not named, by McFarlane and Pilling in 1947.[5] Urokinase was originally isolated from human urine, and it is also present in the blood and in the extracellular matrix of many tissues. The primary physiological substrate of this enzyme is plasminogen, which is an inactive form (zymogen) of the serine protease plasmin. Activation of plasmin triggers a proteolytic cascade that, depending on the physiological environment, participates in thrombolysis or extracellular matrix degradation. This cascade had been involved in vascular diseases and cancer progression.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 20 10 5 2.5 1.25 0.62 0.31 0
OD 2.32 1.7 1.19 0.72 0.46 0.32 0.22 0.07
Corrected OD 2.25 1.63 1.12 0.65 0.39 0.25 0.15 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Mouse uPA spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 86-101 96
Plasma(n=8) 92-105 103





 

| Linearity

Samples were spiked with high concentrations of Mouse uPA and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4)
1 : 2 Range (%) 84-95 88-96
Average (%) 91 93
1 : 4 Range (%) 89-103 87-108
Average (%) 94 98







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
360 / 360
96T
1
460 / 460
Top