Product Name:
Mouse Angiopoietin 1 (ANG-1) ELISA Kit
Cat.#:
JL11963
Regulatory Status:
Alternative:
ANG1; AGP1; AGPT; ANG-I
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.78-50ng/mL
Sensitivity:
0.36ng/mL
Specificity:
This kit recognizes Mouse ANG-1in samples. No significant cross-reactivity or interference between Mouse ANG-1 and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:

Research Area:
Cardiovascular Biology
Product Datasheets
| ANG-1
Angiopoietin 1 is a type of angiopoietin and is encoded by the gene ANGPT1.Angiopoietins are proteins with important roles in vascular development and angiogenesis. All angiopoietins bind with similar affinity to an endothelial cell-specific tyrosine-protein kinase receptor. The protein encoded by this gene is a secreted glycoprotein that activates the receptor by inducing its tyrosine phosphorylation. It plays a critical role in mediating reciprocal interactions between the endothelium and surrounding matrix and mesenchyme. The protein also contributes to blood vessel maturation and stability, and may be involved in early development of the heart.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.Concentration (ng/mL) | 50 | 25 | 12.5 | 6.25 | 3.12 | 1.56 | 0.78 | 0 |
OD | 2.39 | 1.48 | 0.86 | 0.51 | 0.32 | 0.23 | 0.18 | 0.07 |
Corrected OD | 2.32 | 1.41 | 0.79 | 0.44 | 0.25 | 0.16 | 0.11 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Mouse ANG-1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.Sample Type | Range(%) | Average Recovery(%) |
Serum(n=8) | 84-101 | 96 |
Plasma(n=8) | 91-105 | 102 |
Cell Culture Supernates(n=8) | 96-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Mouse ANG-1 and diluted with Universal Diluent to produce samples with values within the range of the assay.Serum(n=4) | Plasma(n=4) | Cell Culture Supernates(n=4) | ||
1 : 2 | Range (%) | 84-95 | 88-96 | 90-110 |
Average (%) | 91 | 93 | 96 | |
1 : 4 | Range (%) | 89-103 | 87-108 | 105-115 |
Average (%) | 94 | 98 | 108 |






Size
Quantity
Price ($)
48T


360 / 360
96T


460 / 460