Product Name:
Human Chorionic Gonadotropin (HCG) ELISA Kit
Cat.#:
JL11726
Regulatory Status:
Alternative:
Chorionic Gonadotropin
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
78.12-5000pg/mL
Sensitivity:
28.76pg/mL
Specificity:
This kit recognizes Human HCG in samples. No significant cross-reactivity or interference between Human HCG and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:

Research Area:
Developmental Biology
Product Datasheets
| HCG
Human chorionic gonadotropin (hCG) is a hormone for the maternal recognition of pregnancy produced by trophoblast cells that are surrounding a growing embryo (syncytiotrophoblast initially), which eventually forms the placenta after implantation.[1][2] The presence of hCG is detected in some pregnancy tests (HCG pregnancy strip tests). Some cancerous tumors produce this hormone; therefore, elevated levels measured when the patient is not pregnant may lead to a cancer diagnosis and, if high enough, paraneoplastic syndromes, however, it is unknown whether this production is a contributing cause or an effect of carcinogenesis.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.Concentration (pg/mL) | 5000 | 2500 | 1250 | 625 | 312.5 | 156.25 | 78.12 | 0 |
OD | 2.59 | 1.57 | 0.94 | 0.64 | 0.38 | 0.26 | 0.23 | 0.08 |
Corrected OD | 2.51 | 1.49 | 0.86 | 0.56 | 0.3 | 0.18 | 0.15 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human HCG spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.Sample Type | Range(%) | Average Recovery(%) |
Serum(n=8) | 84-101 | 96 |
Plasma(n=8) | 92-105 | 102 |
Cell Culture Supernates(n=8) | 96-108 | 104 |
| Linearity
Samples were spiked with high concentrations of Human HCG and diluted with Universal Diluent to produce samples with values within the range of the assay.Serum(n=4) | Plasma(n=4) | Cell Culture Supernates(n=4) | ||
1 : 2 | Range (%) | 84-95 | 88-96 | 90-110 |
Average (%) | 91 | 93 | 96 | |
1 : 4 | Range (%) | 89-103 | 87-108 | 105-115 |
Average (%) | 94 | 98 | 108 |






Size
Quantity
Price ($)
48T


360 / 360
96T


460 / 460