Product Name:
Human Apolipoprotein B (apo-B) ELISA Kit
Cat.#:
JL18373
Regulatory Status:
Alternative:
Apo-B; FLDB; Apo B-100; Apo B-48
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.31-20ug/mL
Sensitivity:
0.11ug/mL
Specificity:
This kit recognizes Human apo-Bin samples. No significant cross-reactivity or interference between Human apo-B and analogues was observed.
Sample Type:
serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:

Research Area:
Metabolism
Product Datasheets
| apo-B
Apolipoprotein B (ApoB) is a protein that in humans is encoded by the APOB gene. Its measurement is commonly used to detect risk of atherosclerotic cardiovascular disease.Apolipoprotein B is the primary apolipoprotein of chylomicrons, VLDL, Lp(a), IDL, and LDL particles (LDL—commonly known as "bad cholesterol" when in reference to both heart disease and vascular disease in general), which is responsible for carrying fat molecules (lipids), including cholesterol, around the body to all cells within all tissues.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.Concentration (ug/mL) | 20 | 10 | 5 | 2.5 | 1.25 | 0.62 | 0.31 | 0 |
OD | 2.24 | 1.63 | 1.01 | 0.55 | 0.44 | 0.31 | 0.25 | 0.09 |
Corrected OD | 2.15 | 1.54 | 0.92 | 0.46 | 0.35 | 0.22 | 0.16 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human apo-B spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.Sample Type | Range(%) | Average Recovery(%) |
Serum(n=8) | 84-101 | 96 |
Plasma(n=8) | 92-105 | 102 |
Cell Culture Supernates(n=8) | 96-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Human apo-B and diluted with Universal Diluent to produce samples with values within the range of the assay.Serum(n=4) | Plasma(n=4) | Cell Culture Supernates(n=4) | ||
1 : 2 | Range (%) | 84-95 | 88-96 | 90-110 |
Average (%) | 91 | 93 | 96 | |
1 : 4 | Range (%) | 89-103 | 87-108 | 105-115 |
Average (%) | 94 | 98 | 108 |






Size
Quantity
Price ($)
48T


360 / 360
96T


460 / 460