400-0066-400

Human Osteocalcin (BGP;OCN) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human Osteocalcin (BGP;OCN) ELISA Kit
Product Name:

Human Osteocalcin (BGP;OCN) ELISA Kit

Cat.#:

JL19437

Regulatory Status:
Alternative:

BGLAP; OT; BGP; Bone Gla Protein; Bone Gamma-Carboxyglutamate Protein; Gamma-carboxyglutamic acid-containing protein

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.625-40ng/mL
Sensitivity:
0.29ng/mL
Specificity:
This kit recognizes Human BGP;OCNin samples. No significant cross-reactivity or interference between Human BGP;OCN and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Metabolism
  • Human Osteocalcin (BGP;OCN) ELISA Kit
  • Human Osteocalcin (BGP;OCN) ELISA Kit
  • Human Osteocalcin (BGP;OCN) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| BGP;OCN

Osteocalcin, also known as bone gamma-carboxyglutamic acid-containing protein (BGLAP), is a small (49-amino-acid) noncollagenous protein hormone found in bone and dentin, first identified as a calcium-binding protein.Because osteocalcin has gla domains, its synthesis is vitamin K dependent. In humans, osteocalcin is encoded by the BGLAP gene.Its receptors include GPRC6A, GPR158, and possibly a third, yet-to-be-identified receptor.There is evidence that GPR37 might be the third osteocalcin receptor.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 40 20 10 5 2.5 1.25 0.625 0
OD 2.16 1.45 0.93 0.56 0.38 0.28 0.21 0.08
Corrected OD 2.08 1.37 0.85 0.48 0.3 0.2 0.13 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human BGP;OCN spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-101 96
Plasma(n=8) 92-105 102
Cell Culture Supernates(n=8) 96-108 105





 

| Linearity

Samples were spiked with high concentrations of Human BGP;OCN and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Culture Supernates(n=4)
1 : 2 Range (%) 84-95 88-96 90-110
Average (%) 91 93 96
1 : 4 Range (%) 89-103 87-108 105-115
Average (%) 94 98 108







 