Product Name:
Human Melanoma Cell Adhesion Molecule (MCAM) ELISA Kit
Cat.#:
JL33739
Regulatory Status:
Alternative:
MCAM; MUC18; Melanoma Cell Adhesion Molecule; Cell surface glycoprotein P1H12; Melanoma-associated antigen A32; S-endo 1 endothelial-associated antigen
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.78-50ng/mL
Sensitivity:
0.21ng/mL
Specificity:
This kit recognizes Human MCAM in samples. No significant cross-reactivity or interference between Human MCAM and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates and other biological fluids
Assay Procedure:

Research Area:
Cell Biology
Product Datasheets
| MCAM
CD146 (cluster of differentiation 146) also known as the melanoma cell adhesion molecule (MCAM) or cell surface glycoprotein MUC18, is a 113kDa cell adhesion molecule currently used as a marker for endothelial cell lineage. The CD146 protein is encoded by the MCAM gene.MCAM functions as a receptor for laminin alpha 4,a matrix molecule that is broadly expressed within the vascular wall.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.Concentration (ng/mL) | 50 | 25 | 12.5 | 6.25 | 3.12 | 1.56 | 0.78 | 0 |
OD | 2.41 | 1.84 | 1.21 | 0.85 | 0.52 | 0.33 | 0.21 | 0.08 |
Corrected OD | 2.33 | 1.76 | 1.13 | 0.77 | 0.44 | 0.25 | 0.13 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human MCAM spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.Sample Type | Range(%) | Average Recovery(%) |
Serum(n=8) | 85-101 | 97 |
Plasma(n=8) | 92-105 | 102 |
Tissue Homogenates(n=8) | 96-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Human MCAM and diluted with Universal Diluent to produce samples with values within the range of the assay.Serum(n=4) | Plasma(n=4) | Tissue Homogenates(n=4) | ||
1 : 2 | Range (%) | 85-95 | 88-96 | 90-110 |
Average (%) | 93 | 92 | 96 | |
1 : 4 | Range (%) | 89-103 | 87-108 | 105-116 |
Average (%) | 94 | 98 | 108 |






Size
Quantity
Price ($)
48T


360 / 360
96T


460 / 460