Product Name:
Human Decorin (DCN) ELISA Kit
Cat.#:
JL14181
Regulatory Status:
Alternative:
CSCD; DSPG2; PG40; PGII; PGS2; PG-S2; SLRR1B; Bone proteoglycan II
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
1.25-80ng/mL
Sensitivity:
0.43ng/mL
Specificity:
This kit recognizes Human DCN in samples. No significant cross-reactivity or interference between Human DCN and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates and other biological fluids
Assay Procedure:

Research Area:
Signal Transduction
Product Datasheets
| DCN
Decorin is a protein that is encoded by the DCN gene.Decorin is a proteoglycan that is on average 90 - 140 kilodaltons (kDa) in molecular weight. It belongs to the small leucine-rich proteoglycan (SLRP) family and consists of a protein core containing leucine repeats with a glycosaminoglycan (GAG) chain consisting of either chondroitin sulfate (CS) or dermatan sulfate (DS).Decorin is a small cellular or pericellular matrix proteoglycan and is closely related in structure to biglycan protein. Decorin and biglycan are thought to be the result of a gene duplication. This protein is a component of connective tissue, binds to type I collagen fibrils, and plays a role in matrix assembly.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.Concentration (ng/mL) | 80 | 40 | 20 | 10 | 5 | 2.5 | 1.25 | 0 |
OD | 2.02 | 1.27 | 0.8 | 0.49 | 0.34 | 0.25 | 0.19 | 0.07 |
Corrected OD | 1.95 | 1.2 | 0.73 | 0.42 | 0.27 | 0.18 | 0.12 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human DCN spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.Sample Type | Range(%) | Average Recovery(%) |
Serum(n=8) | 84-101 | 96 |
Plasma(n=8) | 92-104 | 101 |
Tissue Homogenates(n=8) | 96-108 | 104 |
| Linearity
Samples were spiked with high concentrations of Human DCN and diluted with Universal Diluent to produce samples with values within the range of the assay.Serum(n=4) | Plasma(n=4) | Tissue Homogenates(n=4) | ||
1 : 2 | Range (%) | 85-95 | 88-96 | 90-110 |
Average (%) | 92 | 93 | 96 | |
1 : 4 | Range (%) | 89-103 | 87-108 | 105-115 |
Average (%) | 94 | 98 | 108 |