400-0066-400

Human Epidermal Growth Factor Receptor 2 (Her2) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human Epidermal Growth Factor Receptor 2 (Her2) ELISA Kit
Product Name:

Human Epidermal Growth Factor Receptor 2 (Her2) ELISA Kit

Cat.#:

JL12847

Regulatory Status:
Alternative:

CD340; EGFR2; HER-2/neu;EGFR2; NEU; MLN 19; NGL; TKR1; p185; Epidermal Growth Factor Receptor 2; V-Erb-B2 Erythroblastic Leukemia Viral Oncogene Homolog 2; Neuro/Glioblastoma Derived Oncogene Homolog

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.78-50ng/mL
Sensitivity:
0.25ng/mL
Specificity:
This kit recognizes Human Her2 in samples. No significant cross-reactivity or interference between Human Her2 and analogues was observed.
Sample Type:
serum, plasma, tissue homogenates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Signal Transduction
  • Human Epidermal Growth Factor Receptor 2 (Her2) ELISA Kit
  • Human Epidermal Growth Factor Receptor 2 (Her2) ELISA Kit
  • Human Epidermal Growth Factor Receptor 2 (Her2) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| Her2

Receptor tyrosine-protein kinase erbB-2 is a protein that normally resides in the membranes of cells and is encoded by the ERBB2 gene. ERBB is abbreviated from erythroblastic oncogene B, a gene originally isolated from the avian genome. The human protein is also frequently referred to as HER2 (human epidermal growth factor receptor 2) or CD340 (cluster of differentiation 340).HER2 is a member of the human epidermal growth factor receptor (HER/EGFR/ERBB) family. But contrary to other members of the ERBB family, HER2 does not directly bind ligand. HER2 activation results from heterodimerization with another ERBB member or by homodimerization when HER2 concentration are high, for instance in cancer.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 50 25 12.5 6.25 3.12 1.56 0.78 0
OD 2.03 1.21 0.69 0.41 0.29 0.23 0.19 0.09
Corrected OD 1.94 1.12 0.6 0.32 0.2 0.14 0.1 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human Her2 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-101 96
Plasma(n=8) 91-105 101
Tissue Homogenates(n=8) 96-108 105





 

| Linearity

Samples were spiked with high concentrations of Human Her2 and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Tissue Homogenates(n=4)
1 : 2 Range (%) 84-95 88-96 90-110
Average (%) 91 93 96
1 : 4 Range (%) 89-103 88-108 105-115
Average (%) 94 99 108







 