Product Name:
Human Small Dense Low Density LipoProtein (sdLDL-C) ELISA Kit
Cat.#:
JL19992
Regulatory Status:
Alternative:
sd-LDL-C
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
1.56-100ng/mL
Sensitivity:
0.72ng/mL
Specificity:
This kit recognizes Human sdLDL-C in samples. No significant cross-reactivity or interference between Human sdLDL-C and analogues was observed.
Sample Type:
Serum;plasma;tissue homogenates;cell lysates;cell culture supernates and other biological fluids
Assay Procedure:
Research Area:
Metabolism
Product Datasheets
| sdLDL-C
Small and dense low-density lipoprotein (sdLDL) is a small, dense subtype of low-density lipoprotein (LDL) with a stronger atherosclerotic effect and is an important risk marker for cardiovascular disease 13. Its characteristics include easy oxidation, slow clearance and high affinity for the arterial wall, resulting in foam cell formation and plaque growth.Low levels of antioxidants (such as coenzyme Q10, vitamin E) in sdLDL granules are easily oxidized to oxidized LDL, which promotes foam cell formation.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (ng/mL) | 100 | 50 | 25 | 12.5 | 6.25 | 3.12 | 1.56 | 0 |
| OD | 2.69 | 1.89 | 1.32 | 0.8 | 0.47 | 0.31 | 0.16 | 0.1 |
| Corrected OD | 2.59 | 1.79 | 1.22 | 0.7 | 0.37 | 0.21 | 0.06 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human sdLDL-C spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 84-101 | 96 |
| Plasma(n=8) | 92-106 | 103 |
| Cell Culture Supernates(n=8) | 96-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Human sdLDL-C and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | Cell Culture Supernates(n=4) | ||
| 1 : 2 | Range (%) | 84-95 | 88-95 | 90-110 |
| Average (%) | 90 | 93 | 96 | |
| 1 : 4 | Range (%) | 89-103 | 87-108 | 105-114 |
| Average (%) | 94 | 98 | 107 |
Size
Quantity
Price ($)
48T
1
460 / 460
96T
1
760 / 760