Product Name:
Human Secondary Lymphoid Tissue Chemokine (SLC;CCL21) ELISA Kit
Cat.#:
JL12003
Regulatory Status:
Alternative:
CCL21; 6Ckine; CKb9; ECL; SCYA21; TCA4; ECL; Chemokine(C-C-Motif)Ligand 21; Beta Chemokine Exodus-2; Efficient Chemoattractant For Lymphocytes
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
62.5-4000pg/mL
Sensitivity:
29.5pg/mL
Specificity:
This kit recognizes Human SLC;CCL21 in samples. No significant cross-reactivity or interference between Human SLC;CCL21 and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:

Research Area:
Immunology
Product Datasheets
| SLC;CCL21
Chemokine (C-C motif) ligand 21 (CCL21) is a small cytokine belonging to the CC chemokine family. This chemokine is also known as 6Ckine (because it has six conserved cysteine residues instead of the four cysteines typical to chemokines), exodus-2, and secondary lymphoid-tissue chemokine (SLC).CCL21 elicits its effects by binding to a cell surface chemokine receptor known as CCR7.The main function of CCL21 is to guide CCR7 expressing leukocytes to the secondary lymphoid organs, such as lymph nodes and Peyer´s patches.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.Concentration (pg/mL) | 4000 | 2000 | 1000 | 500 | 250 | 125 | 62.5 | 0 |
OD | 2.04 | 1.61 | 1.23 | 0.84 | 0.53 | 0.35 | 0.23 | 0.08 |
Corrected OD | 1.96 | 1.53 | 1.15 | 0.76 | 0.45 | 0.27 | 0.15 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human SLC;CCL21 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.Sample Type | Range(%) | Average Recovery(%) |
Serum(n=8) | 84-101 | 96 |
Plasma(n=8) | 92-105 | 102 |
Cell Culture Supernates(n=8) | 96-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Human SLC;CCL21 and diluted with Universal Diluent to produce samples with values within the range of the assay.Serum(n=4) | Plasma(n=4) | Cell Culture Supernates(n=4) | ||
1 : 2 | Range (%) | 85-95 | 88-96 | 90-110 |
Average (%) | 92 | 93 | 96 | |
1 : 4 | Range (%) | 89-103 | 87-107 | 105-115 |
Average (%) | 94 | 97 | 108 |






Size
Quantity
Price ($)
48T


360 / 360
96T