Product Name:
Human Kielin/chordin-like Protein (KCP) ELISA Kit
Cat.#:
JL23826
Regulatory Status:
Alternative:
Kielin/chordin-like protein
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.312-20ng/mL
Sensitivity:
0.154ng/mL
Specificity:
This kit recognizes Human KCP in samples. No significant cross-reactivity or interference between Human KCP and analogues was observed.
Sample Type:
Serum;plasma and tissue homogenates
Assay Procedure:
Research Area:
Signal Transduction
Product Datasheets
| KCP
Kielin/Chordin-like protein (KCP) is a novel enhancer of BMP signaling that ameliorates renal fibrotic diseases. Bone morphogenetic proteins (BMPs) critically regulate embryonic development, cell differentiation, and disease processes. BMP signaling is normally inhibited by cysteine-rich proteins such as chordin, which sequester BMP ligands from their receptors. In contrast to these inhibitors, KCP uniquely potentiates BMP signaling through its distinct structural domains, thereby demonstrating therapeutic potential for fibrotic disorders by promoting tissue repair pathways while suppressing excessive extracellular matrix deposition characteristic of fibrosis.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (ng/mL) | 20 | 10 | 5 | 2.5 | 1.25 | 0.625 | 0.312 | 0 |
| OD | 2.45 | 1.88 | 1.46 | 1.13 | 0.76 | 0.49 | 0.38 | 0.08 |
| Corrected OD | 2.37 | 1.8 | 1.38 | 1.05 | 0.68 | 0.41 | 0.3 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human KCP spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 84-101 | 96 |
| Plasma(n=8) | 92-105 | 102 |
| Linearity
Samples were spiked with high concentrations of Human KCP and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | ||
| 1 : 2 | Range (%) | 84-95 | 88-96 |
| Average (%) | 91 | 93 | |
| 1 : 4 | Range (%) | 89-104 | 87-108 |
| Average (%) | 94 | 98 |
Size
Quantity
Price ($)
48T
1
280 / 280
96T
1
460 / 460