Product Name:
Human NANOG Homeobox Protein (NANOG) ELISA Kit
Cat.#:
JL19865
Regulatory Status:
Alternative:
Homeobox transcription factor Nanog
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.156-10ng/mL
Sensitivity:
0.074ng/mL
Specificity:
This kit recognizes Human NANOG in samples. No significant cross-reactivity or interference between Human NANOG and analogues was observed.
Sample Type:
Tissue homogenates;cell lysates and other biological fluids
Assay Procedure:
Research Area:
Stem Cells
Product Datasheets
| NANOG
NANOG Homeobox protein is a transcriptional factor that helps embryonic stem cells (ESCs) maintain pluripotency by suppressing cell determination factors.NANOG is fundamental for self-renewal and pluripotency, and NANOG p8 is highly expressed in cancer cells.NANOG is highly expressed in cancer stem cells and may thus function as an oncogene to promote carcinogenesis.NANOG p8 gene has been evaluated as a prognostic and predictive cancer biomarker.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (ng/mL) | 10 | 5 | 2.5 | 1.25 | 0.625 | 0.312 | 0.156 | 0 |
| OD | 2.38 | 1.47 | 0.84 | 0.48 | 0.32 | 0.25 | 0.2 | 0.09 |
| Corrected OD | 2.29 | 1.38 | 0.75 | 0.39 | 0.23 | 0.16 | 0.11 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human NANOG spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Tissue Homogenates(n=8) | 84-101 | 96 |
| Cell Lysates(n=8) | 92-105 | 101 |
| Linearity
Samples were spiked with high concentrations of Human NANOG and diluted with Universal Diluent to produce samples with values within the range of the assay.| Tissue Homogenates(n=4) | Cell Lysates(n=4) | ||
| 1 : 2 | Range (%) | 88-96 | 90-110 |
| Average (%) | 93 | 96 | |
| 1 : 4 | Range (%) | 87-108 | 105-115 |
| Average (%) | 98 | 108 |
Size
Quantity
Price ($)
48T
1
360 / 360
96T
1
460 / 460