400-0066-400

Human P-Selectin GlycoProtein Ligand 1 (PSGL1) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human P-Selectin GlycoProtein Ligand 1 (PSGL1) ELISA Kit
Product Name:

Human P-Selectin GlycoProtein Ligand 1 (PSGL1) ELISA Kit

Cat.#:

JL14376

Regulatory Status:
Alternative:

CD162; SELPLG; CLA; Selectin P Ligand

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.625-40ng/mL
Sensitivity:
0.2ng/mL
Specificity:
This kit recognizes Human PSGL1 in samples. No significant cross-reactivity or interference between Human PSGL1 and analogues was observed.
Sample Type:
Serum;plasma;tissue homogenates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Cell Biology
  • Human P-Selectin GlycoProtein Ligand 1 (PSGL1) ELISA Kit
  • Human P-Selectin GlycoProtein Ligand 1 (PSGL1) ELISA Kit
  • Human P-Selectin GlycoProtein Ligand 1 (PSGL1) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| PSGL1

P-Selectin GlycoProtein Ligand 1 (PSGL1) , the high affinity counter-receptor for P-selectin on myeloid cells and stimulated T lymphocytes.PSGL-1 is expressed on all white blood cells and plays an important role in the recruitment of white blood cells into inflamed tissue: White blood cells normally do not interact with the endothelium of blood vessels. However, inflammation causes the expression of cell adhesion molecules (CAM) such as P-selectin on the surface of the blood vessel wall. White blood cells present in flowing blood can interact with CAM. The first step in this interaction process is carried out by PSGL-1 interacting with P-selectin and/or E-selectin on endothelial cells and adherent platelets.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 40 20 10 5 2.5 1.25 0.625 0
OD 2.52 1.96 1.38 0.97 0.55 0.31 0.19 0.08
Corrected OD 2.44 1.88 1.3 0.88 0.47 0.23 0.11 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human PSGL1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-101 96
Plasma(n=8) 92-105 102
Tissue Homogenates(n=8) 92-105 102





 

| Linearity

Samples were spiked with high concentrations of Human PSGL1 and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Tissue Homogenates(n=4)
1 : 2 Range (%) 84-95 88-96 90-110
Average (%) 91 93 96
1 : 4 Range (%) 89-103 87-108 87-108
Average (%) 94 98 98







 
Close