Product Name:
Human Thrombin/antithrombin Complex (TAT) ELISA Kit
Cat.#:
JL12851
Regulatory Status:
Alternative:
TAT
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
6.25-400pg/mL
Sensitivity:
3.16pg/mL
Specificity:
This kit recognizes Human TAT in samples. No significant cross-reactivity or interference between Human TAT and analogues was observed.
Sample Type:
Plasma
Assay Procedure:
Research Area:
Cardiovascular Biology
Product Datasheets
| TAT
Thrombin–antithrombin complex (TAT) is a protein complex of thrombin and antithrombin.It is a marker of net activation of coagulation.TAT is formed in response to the high thrombin level caused by coagulation following a ruptured vessel. Since thrombin is rapidly bound by antithrombin, TAT is a useful measure for thrombin level in the blood. Thrombin can pass the blood–brain barrier, destroying neurons and potentially causing cerebral edemas.The half-life of TAT is approximately 15 minutes.TAT levels are increased with pregnancy and by ethinylestradiol-containing birth control pills.They have also been reported to be increased by menopausal hormone therapy, although findings are mixed, and with high-dose parenteral estradiol therapy for prostate cancer.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (pg/mL) | 400 | 200 | 100 | 50 | 25 | 12.5 | 6.25 | 0 |
| OD | 2.22 | 1.56 | 0.92 | 0.63 | 0.42 | 0.25 | 0.18 | 0.08 |
| Corrected OD | 2.14 | 1.48 | 0.84 | 0.55 | 0.34 | 0.17 | 0.1 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human TAT spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Plasma(n=8) | 92-105 | 103 |
| Linearity
Samples were spiked with high concentrations of Human TAT and diluted with Universal Diluent to produce samples with values within the range of the assay.| Plasma(n=4) | ||
| 1 : 2 | Range (%) | 88-95 |
| Average (%) | 93 | |
| 1 : 4 | Range (%) | 87-108 |
| Average (%) | 97 |
Size
Quantity
Price ($)
48T
1
420 / 420
96T
1
640 / 640


