400-0066-400

Human Endothelial Cell Specific Molecule 1 (ESM1) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human Endothelial Cell Specific Molecule 1 (ESM1) ELISA Kit
Product Name:

Human Endothelial Cell Specific Molecule 1 (ESM1) ELISA Kit

Cat.#:

JL19689

Regulatory Status:
Alternative:

Endocan

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.15-10ng/mL
Sensitivity:
0.04ng/mL
Specificity:
This kit recognizes Human ESM1 in samples. No significant cross-reactivity or interference between Human ESM1 and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Metabolism
  • Human Endothelial Cell Specific Molecule 1 (ESM1) ELISA Kit
  • Human Endothelial Cell Specific Molecule 1 (ESM1) ELISA Kit
  • Human Endothelial Cell Specific Molecule 1 (ESM1) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| ESM1

Endothelial cell-specific molecule 1 is a protein that in humans is encoded by the ESM1 gene.This gene encodes a secreted protein which is mainly expressed in the endothelial cells in human lung and kidney tissues. The expression of this gene is regulated by cytokines, suggesting that it may play a role in endothelium-dependent pathological disorders. The transcript contains multiple polyadenylation and mRNA instability signals.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 10 5 2.5 1.25 0.62 0.31 0.15 0
OD 2.39 1.67 1.17 0.71 0.52 0.33 0.27 0.1
Corrected OD 2.29 1.57 1.07 0.61 0.42 0.23 0.17 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human ESM1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-102 96
Plasma(n=8) 93-105 102
Cell Culture Supernates(n=8) 96-108 105





 

| Linearity

Samples were spiked with high concentrations of Human ESM1 and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Culture Supernates(n=4)
1 : 2 Range (%) 84-95 88-96 90-110
Average (%) 91 93 96
1 : 4 Range (%) 89-103 87-108 105-115
Average (%) 94 98 108







 