400-0066-400

Human Microalbunminuria (MAU;ALB) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human Microalbunminuria (MAU;ALB) ELISA Kit
Product Name:

Human Microalbunminuria (MAU;ALB) ELISA Kit

Cat.#:

JL15013

Regulatory Status:
Alternative:

mau;alb

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.78-50ng/mL
Sensitivity:
0.39ng/mL
Specificity:
This kit recognizes Human mau;alb in samples. No significant cross-reactivity or interference between Human mau;alb and analogues was observed.
Sample Type:
Serum;plasma;tissue homogenates;cell lysates;cell culture supernates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Metabolism
  • Human Microalbunminuria (MAU;ALB) ELISA Kit
  • Human Microalbunminuria (MAU;ALB) ELISA Kit
  • Human Microalbunminuria (MAU;ALB) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| mau;alb

Microalbuminuria (MAU) refers to the presence of small amounts of albumin in urine that are detectable by sensitive assays but not by conventional dipstick tests. The glomerular capillary basement membrane normally functions as a selective filter with pore diameters of approximately 5.5 nm. Since albumin molecules circulating in blood have a diameter of 7.2 nm, healthy kidneys prevent their passage into urine through both size-selective and charge-selective barriers.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 50 25 12.5 6.25 3.12 1.56 0.78 0
OD 2.47 1.73 1.33 0.92 0.63 0.44 0.32 0.09
Corrected OD 2.38 1.64 1.24 0.83 0.54 0.35 0.23 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human mau;alb spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-101 96
Plasma(n=8) 92-105 102
Cell Culture Supernates(n=8) 96-108 105





 

| Linearity

Samples were spiked with high concentrations of Human mau;alb and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Culture Supernates(n=4)
1 : 2 Range (%) 84-95 88-96 90-110
Average (%) 91 93 96
1 : 4 Range (%) 89-103 87-108 105-115
Average (%) 94 98 108







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
360 / 360
96T
1
460 / 460
Top