400-0066-400

Human Angiostatin (angiostatin) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human Angiostatin (angiostatin) ELISA Kit
Product Name:

Human Angiostatin (angiostatin) ELISA Kit

Cat.#:

JL11881

Regulatory Status:
Alternative:

Angiostatin

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
62.5-4000pg/mL
Sensitivity:
23.51pg/mL
Specificity:
This kit recognizes Human angiostatinin samples. No significant cross-reactivity or interference between Human angiostatin and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Cancer
  • Human Angiostatin (angiostatin) ELISA Kit
  • Human Angiostatin (angiostatin) ELISA Kit
  • Human Angiostatin (angiostatin) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| angiostatin

Angiostatin is a naturally occurring protein found in several animal species, including humans. It is an endogenous angiogenesis inhibitor (i.e., it blocks the growth of new blood vessels). Clinical trials have been undertaken for its use in anticancer therapy.Angiostatin is a 38 kDa fragment of a larger protein, plasmin (itself a fragment of plasminogen) enclosing three to five contiguous kringle modules. Each module contains two small beta sheets and three disulfide bonds.There are four different structural variants to angiostatin differing in the combination of kringle domains: K1-3, K1-4, K1-5, K1-4 with a fragment of K-5. Each kringle domain contributes a different element of inhibition to the cytokine.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (pg/mL) 4000 2000 1000 500 250 125 62.5 0
OD 2.45 1.6 0.94 0.6 0.39 0.27 0.24 0.09
Corrected OD 2.36 1.51 0.85 0.51 0.3 0.18 0.15 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human angiostatin spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Tissue Homogenates(n=8) 84-101 95
Cell Lysates(n=8) 92-105 102





 

| Linearity

Samples were spiked with high concentrations of Human angiostatin and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Tissue Homogenates(n=4) Cell Lysates(n=4)
1 : 2 Range (%) 84-95 88-96
Average (%) 91 93
1 : 4 Range (%) 89-103 87-108
Average (%) 94 98







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
360 / 360
96T
1
460 / 460
Top