Product Name:
Human Protein DJ-1 (DJ-1) ELISA Kit
Cat.#:
JL14590
Regulatory Status:
Alternative:
DJ1; Autosomal Recessive;Early Onset
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
1.56-100ng/mL
Sensitivity:
0.79ng/mL
Specificity:
This kit recognizes Human DJ-1 in samples. No significant cross-reactivity or interference between Human DJ-1 and analogues was observed.
Sample Type:
Serum;plasma;tissue homogenates;cell lysates and other biological fluids
Assay Procedure:
Research Area:
Neuroscience
Product Datasheets
| DJ-1
DJ1, also known as Parkinson disease protein 7, is a protein which is encoded by the PARK7 gene. Its weak glyoxalase activity has been verified by many labs, however the reported protein deglycase activity is likely to be an artifact stemming from DJ-1's ability to destroy free methylglyoxal.Under an oxidative condition, DJ-1 inhibits the aggregation of α-synuclein via its chaperone activity, thus functioning as a redox-sensitive chaperone and as a sensor for oxidative stress. Accordingly, DJ-1 apparently protects neurons against oxidative stress and cell death.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (ng/mL) | 100 | 50 | 25 | 12.5 | 6.25 | 3.12 | 1.56 | 0 |
| OD | 2.36 | 1.78 | 1.23 | 0.79 | 0.46 | 0.29 | 0.17 | 0.09 |
| Corrected OD | 2.27 | 1.69 | 1.14 | 0.7 | 0.37 | 0.2 | 0.08 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human DJ-1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 84-102 | 97 |
| Plasma(n=8) | 92-105 | 102 |
| Tissue Homogenates(n=8) | 96-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Human DJ-1 and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | Tissue Homogenates(n=4) | ||
| 1 : 2 | Range (%) | 84-95 | 88-96 | 90-110 |
| Average (%) | 91 | 93 | 96 | |
| 1 : 4 | Range (%) | 89-102 | 87-109 | 105-115 |
| Average (%) | 93 | 98 | 108 |
Size
Quantity
Price ($)
48T
1
360 / 360
96T
1
460 / 460