400-0066-400

Human Amyloid-β (Aβ) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human Amyloid-β (Aβ) ELISA Kit
Product Name:

Human Amyloid-β (Aβ) ELISA Kit

Cat.#:

JL14446

Regulatory Status:
Alternative:

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
15.62-1000pg/mL
Sensitivity:
7.82pg/mL
Specificity:
This kit recognizes Human Aβ in samples. No significant cross-reactivity or interference between Human Aβ and analogues was observed.
Sample Type:
Serum;plasma;tissue homogenates;cell lysates;cell culture supernates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Neuroscience
  • Human Amyloid-β (Aβ) ELISA Kit
  • Human Amyloid-β (Aβ) ELISA Kit
  • Human Amyloid-β (Aβ) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| 

Amyloid beta (Aβ, Abeta or beta-amyloid) denotes peptides of 36–43 amino acids that are the main component of the amyloid plaques found in the brains of people with Alzheimer's disease.The peptides derive from the amyloid-beta precursor protein (APP), which is cleaved by beta secretase and gamma secretase to yield Aβ in a cholesterol-dependent process and substrate presentation. Both neurons and oligodendrocytes produce and release Aβ in the brain, contributing to formation of amyloid plaques.Aβ molecules can aggregate to form flexible soluble oligomers which may exist in several forms.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (pg/mL) 1000 500 250 125 62.5 31.25 15.62 0
OD 2.16 1.45 0.93 0.56 0.38 0.28 0.21 0.08
Corrected OD 2.08 1.37 0.85 0.48 0.3 0.2 0.13 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human Aβ spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-101 96
Plasma(n=8) 92-105 102
Cell Culture Supernates(n=8) 96-108 105





 

| Linearity

Samples were spiked with high concentrations of Human Aβ and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Culture Supernates(n=4)
1 : 2 Range (%) 84-96 88-96 90-110
Average (%) 91 93 96
1 : 4 Range (%) 89-103 87-108 105-115
Average (%) 94 98 108







 