400-0066-400

Human C Reactive Protein (CRP) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human C Reactive Protein (CRP) ELISA Kit
Product Name:

Human C Reactive Protein (CRP) ELISA Kit

Cat.#:

JL13865

Regulatory Status:
Alternative:

C-RP; PTX1; Pentraxin-Related; Pentraxin 1

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
31.25-2000pg/mL
Sensitivity:
15.9pg/mL
Specificity:
This kit recognizes Human CRP in samples. No significant cross-reactivity or interference between Human CRP and analogues was observed.
Sample Type:
serum, plasma, tissue homogenates, cell lysates, urine, cerebrospinal fluid, cell culture supernates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Immunology
  • Human C Reactive Protein (CRP) ELISA Kit
  • Human C Reactive Protein (CRP) ELISA Kit
  • Human C Reactive Protein (CRP) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| CRP

C-reactive protein (CRP) is an annular (ring-shaped) pentameric protein found in blood plasma, whose circulating concentrations rise in response to inflammation. It is an acute-phase protein of hepatic origin that increases following interleukin-6 secretion by macrophages and T cells. Its physiological role is to bind to lysophosphatidylcholine expressed on the surface of dead or dying cells (and some types of bacteria) in order to activate the complement system via C1q.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (pg/mL) 2000 1000 500 250 125 62.5 31.25 0
OD 2.02 1.4 0.86 0.59 0.38 0.26 0.18 0.08
Corrected OD 1.94 1.32 0.78 0.51 0.3 0.18 0.1 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human CRP spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-102 96
Plasma(n=8) 92-106 102
Cell Culture Supernates(n=8) 96-108 105





 

| Linearity

Samples were spiked with high concentrations of Human CRP and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Culture Supernates(n=4)
1 : 2 Range (%) 84-95 88-96 90-111
Average (%) 91 93 96
1 : 4 Range (%) 89-103 86-108 105-115
Average (%) 94 98 108







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
280 / 280
96T
1
360 / 360