400-0066-400

Human Interleukin 8 (IL-8) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human Interleukin 8 (IL-8) ELISA Kit
Product Name:

Human Interleukin 8 (IL-8) ELISA Kit

Cat.#:

JL19291

Regulatory Status:
Alternative:

CXCL8; AMCF-I; GCP1; K60; LECT; LUCT; LYNAP; MDNCF; MONAP; NAF; NAP1; SCYB8; TSG1; B-ENAP; Neutrophil-Activating Protein 1; Granulocyte Chemotactic Protein 1

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
7.8-500pg/mL
Sensitivity:
3.4pg/mL
Specificity:
This kit recognizes Human IL-8 in samples. No significant cross-reactivity or interference between Human IL-8 and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Immunology
  • Human Interleukin 8 (IL-8) ELISA Kit
  • Human Interleukin 8 (IL-8) ELISA Kit
  • Human Interleukin 8 (IL-8) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| IL-8

Interleukin 8 (IL-8 or chemokine (C-X-C motif) ligand 8, CXCL8) is a chemokine produced by macrophages and other cell types such as epithelial cells, airway smooth muscle cellsand endothelial cells. Endothelial cells store IL-8 in their storage vesicles, the Weibel–Palade bodies.The interleukin-8 protein is encoded by the CXCL8 gene.IL-8 is initially produced as a precursor peptide of 99 amino acids which then undergoes cleavage to create several active IL-8 isoforms.In culture, a 72 amino acid peptide is the major form secreted by macrophages.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (pg/mL) 500 250 125 62.5 31.2 15.6 7.8 0
OD 2.51 1.76 1.14 0.72 0.46 0.3 0.22 0.09
Corrected OD 2.42 1.67 1.05 0.63 0.37 0.21 0.13 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human IL-8 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-101 96
Plasma(n=8) 92-105 102
Cell Culture Supernates(n=8) 96-108 105





 

| Linearity

Samples were spiked with high concentrations of Human IL-8 and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Culture Supernates(n=4)
1 : 2 Range (%) 84-95 88-96 90-110
Average (%) 91 93 96
1 : 4 Range (%) 89-103 87-108 105-115
Average (%) 94 98 108







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
280 / 280
96T
1
360 / 360
Top