400-0066-400

Human Stromal Cell Derived Factor 1 (SDF-1a;CXCL12) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human Stromal Cell Derived Factor 1 (SDF-1a;CXCL12) ELISA Kit
Product Name:

Human Stromal Cell Derived Factor 1 (SDF-1a;CXCL12) ELISA Kit

Cat.#:

JL34680

Regulatory Status:
Alternative:

SDF-1a;CXCL12

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.15-10ng/mL
Sensitivity:
0.077ng/mL
Specificity:
This kit recognizes Human SDF-1a;CXCL12 in samples. No significant cross-reactivity or interference between Human SDF-1a;CXCL12 and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Cancer
  • Human Stromal Cell Derived Factor 1 (SDF-1a;CXCL12) ELISA Kit
  • Human Stromal Cell Derived Factor 1 (SDF-1a;CXCL12) ELISA Kit
  • Human Stromal Cell Derived Factor 1 (SDF-1a;CXCL12) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| SDF-1a;CXCL12

This assay employs the sandwich enzyme immunoassay technique. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human SDF-1a;CXCL12. Add Sample, Standard, and Streptavidin-HRP by turn, with incubation and washing. TMB Substrate is added for color development, and TMB will appear blue in color catalyzed by HRP and to a final yellow color in reaction to acid. The intensity of the color is positive to the Human SDF-1a;CXCL12 in the sample. The optical density (OD) is measured spectrophotometrically at a wavelength of 450nm. You can calculate the concentration of Human SDF-1a;CXCL12 in the samples by comparing the OD of the samples to the standard curve.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 10 5 2.5 1.25 0.62 0.31 0.15 0
OD 2.48 1.65 1.07 0.71 0.45 0.28 0.19 0.09
Corrected OD 2.39 1.56 0.98 0.62 0.36 0.19 0.1 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human SDF-1a;CXCL12 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-101 96
Plasma(n=8) 92-105 102
Cell Culture Supernates(n=8) 96-108 105





 

| Linearity

Samples were spiked with high concentrations of Human SDF-1a;CXCL12 and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Culture Supernates(n=4)
1 : 2 Range (%) 84-95 88-95 90-110
Average (%) 91 92 96
1 : 4 Range (%) 89-103 87-108 105-115
Average (%) 94 98 108







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
400 / 400
96T
1
560 / 560
Top