400-0066-400

Human Syntenin 2 (ST2) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human Syntenin 2 (ST2) ELISA Kit
Product Name:

Human Syntenin 2 (ST2) ELISA Kit

Cat.#:

JL15409

Regulatory Status:
Alternative:

SDCBP2; SITAC18; Syndecan Binding Protein 2

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.78-50ng/mL
Sensitivity:
0.37ng/mL
Specificity:
This kit recognizes Human ST2 in samples. No significant cross-reactivity or interference between Human ST2 and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Cancer
  • Human Syntenin 2 (ST2) ELISA Kit
  • Human Syntenin 2 (ST2) ELISA Kit
  • Human Syntenin 2 (ST2) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| ST2

Syntenin-2 is a protein that is encoded by the SDCBP2 gene.The protein encoded by this gene contains two class II PDZ domains. PDZ domains facilitate protein-protein interactions by binding to the cytoplasmic C-terminus of transmembrane proteins, and PDZ-containing proteins mediate cell signaling and the organization of protein complexes. The encoded protein binds to phosphatidylinositol 4, 5-bisphosphate (PIP2) and plays a role in nuclear PIP2 organization and cell division. Alternatively spliced transcript variants encoding multiple isoforms have been observed for this gene.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 50 25 12.5 6.25 3.12 1.56 0.78 0
OD 2.64 1.78 1.09 0.64 0.41 0.31 0.23 0.09
Corrected OD 2.55 1.69 1 0.55 0.32 0.22 0.14 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human ST2 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 89-101 96
Plasma(n=8) 92-105 102
Tissue Homogenates(n=8) 96-108 102





 

| Linearity

Samples were spiked with high concentrations of Human ST2 and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Tissue Homogenates(n=4)
1 : 2 Range (%) 84-95 88-96 90-110
Average (%) 91 93 96
1 : 4 Range (%) 89-103 87-108 105-115
Average (%) 94 98 109







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
360 / 360
96T
1
460 / 460
Top