Product Name:
Human Mannose Binding Lectin (MBP;MBL) ELISA Kit
Cat.#:
JL12063
Regulatory Status:
Alternative:
MBL2; COLEC1; HSMBPC; MBP1; MBP; Collectin-1; Mannose-binding protein C; Mannan Binding Protein; Mannose-Binding Lectin(Protein C)2,Soluble(Opsonic Defect)
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.62-40ng/mL
Sensitivity:
0.26ng/mL
Specificity:
This kit recognizes Human MBP;MBLin samples. No significant cross-reactivity or interference between Human MBP;MBL and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:

Research Area:
Signal Transduction
Product Datasheets
| MBP;MBL
Mannose-binding lectin (MBL), also called mannan-binding lectin or mannan-binding protein (MBP), is a lectin that is instrumental in innate immunity as an opsonin and via the lectin pathway.MBL has an oligomeric structure (400-700 kDa), built of subunits that contain three presumably identical peptide chains of about 30 kDa each.Although MBL can form several oligomeric forms, there are indications that dimers and trimers are biologically inactive as an opsonin and at least a tetramer form is needed for activation of complement.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.Concentration (ng/mL) | 40 | 20 | 10 | 5 | 2.5 | 1.25 | 0.62 | 0 |
OD | 2.21 | 1.48 | 1.03 | 0.69 | 0.41 | 0.3 | 0.19 | 0.05 |
Corrected OD | 2.16 | 1.43 | 0.98 | 0.64 | 0.36 | 0.25 | 0.14 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human MBP;MBL spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.Sample Type | Range(%) | Average Recovery(%) |
Serum(n=8) | 84-101 | 96 |
Plasma(n=8) | 92-105 | 102 |
Cell Culture Supernates(n=8) | 96-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Human MBP;MBL and diluted with Universal Diluent to produce samples with values within the range of the assay.Serum(n=4) | Plasma(n=4) | Cell Culture Supernates(n=4) | ||
1 : 2 | Range (%) | 84-95 | 88-96 | 90-110 |
Average (%) | 91 | 93 | 96 | |
1 : 4 | Range (%) | 89-103 | 87-108 | 105-115 |
Average (%) | 94 | 98 | 109 |