400-0066-400

Human Pancreatic Elastase 1 (ELA1) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human Pancreatic Elastase 1 (ELA1) ELISA Kit
Product Name:

Human Pancreatic Elastase 1 (ELA1) ELISA Kit

Cat.#:

JL31461

Regulatory Status:
Alternative:

PE; CELA1; Chymotrypsin-Like Elastase Family,Member 1

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
31.2-2000pg/mL
Sensitivity:
15.4pg/mL
Specificity:
This kit recognizes Human ELA1 in samples. No significant cross-reactivity or interference between Human ELA1 and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Cancer
  • Human Pancreatic Elastase 1 (ELA1) ELISA Kit
  • Human Pancreatic Elastase 1 (ELA1) ELISA Kit
  • Human Pancreatic Elastase 1 (ELA1) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| ELA1

Pancreatic elastase 1 is a serine endopeptidase, a specific type of protease that has the amino acid serine at its active site. Although the recommended name is pancreatic elastase, it can also be referred to as elastase-1, pancreatopeptidase, PE, or serine elastase.Pancreatic elastase 1, was initially thought to be expressed in the pancreas. However it was later discovered that it was the only chymotrypsin-like elastase that was not expressed in the pancreas. In fact, pancreatic elastase is expressed in basal layers of epidermis (at protein level). Hence pancreatic elastase 1 has been renamed elastase 1 (ELA1) or chymotrypsin-like elastase family, member 1 (CELA1).
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (pg/mL) 2000 1000 500 250 125 62.5 31.2 0
OD 2.45 1.88 1.46 1.13 0.76 0.49 0.38 0.08
Corrected OD 2.37 1.8 1.38 1.05 0.68 0.41 0.3 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human ELA1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-101 96
Plasma(n=8) 92-105 102
Cell Lysates(n=8) 96-108 105





 

| Linearity

Samples were spiked with high concentrations of Human ELA1 and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Lysates(n=4)
1 : 2 Range (%) 85-95 88-96 90-110
Average (%) 91 93 96
1 : 4 Range (%) 89-103 87-108 105-115
Average (%) 94 98 108







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
420 / 420
96T
1
640 / 640
Top