Product Name:
Human Biglycan (BGN) ELISA Kit
Cat.#:
JL52544
Regulatory Status:
Alternative:
DSPG1; PG-S1; PGI; SLRR1A; Biglycan Proteoglycan; Bone/cartilage proteoglycan I
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.78-50ng/mL
Sensitivity:
0.21ng/mL
Specificity:
This kit recognizes Human BGN in samples. No significant cross-reactivity or interference between Human BGN and analogues was observed.
Sample Type:
Serum, plasma and other biological fluids
Assay Procedure:

Research Area:
Signal Transduction
Product Datasheets
| BGN
Biglycan is a small leucine-rich repeat proteoglycan (SLRP) which is found in a variety of extracellular matrix tissues, including bone, cartilage and tendon. Biglycan is encoded by the BGN gene which is located on the X chromosome.Biglycan is believed to play a role in the mineralization of bone. Knock-out mice that have had the gene for biglycan suppressed (Bgn -/-) have an osteoporosis-like phenotype with reduced growth rate and lower bone mass than mice that can express biglycan.Biglycan core protein binds to the growth factors BMP-4 and influences its bioactivity. It has also been reported that the presence of biglycan is necessary for BMP-4 to exert its effects on osteoblasts.There is also evidence that biglycan binds to TGF-beta 1.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.Concentration (ng/mL) | 50 | 25 | 12.5 | 6.25 | 3.12 | 1.56 | 0.78 | 0 |
OD | 2.59 | 1.95 | 1.46 | 0.94 | 0.63 | 0.43 | 0.32 | 0.1 |
Corrected OD | 2.49 | 1.85 | 1.36 | 0.84 | 0.53 | 0.33 | 0.22 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human BGN spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.Sample Type | Range(%) | Average Recovery(%) |
Serum(n=8) | 84-102 | 96 |
Plasma(n=8) | 92-106 | 103 |
| Linearity
Samples were spiked with high concentrations of Human BGN and diluted with Universal Diluent to produce samples with values within the range of the assay.Serum(n=4) | Plasma(n=4) | ||
1 : 2 | Range (%) | 84-95 | 88-96 |
Average (%) | 91 | 93 | |
1 : 4 | Range (%) | 89-103 | 87-107 |
Average (%) | 94 | 98 |






Size
Quantity
Price ($)
48T


360 / 360
96T


460 / 460