400-0066-400

Human TEK Tyrosine Kinase, Endothelial (Tie2) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human TEK Tyrosine Kinase, Endothelial (Tie2) ELISA Kit
Product Name:

Human TEK Tyrosine Kinase, Endothelial (Tie2) ELISA Kit

Cat.#:

JL14747

Regulatory Status:
Alternative:

CD202b; CD202-B; TEK; VMCM1; Tyrosine Kinase With Immunoglobulin Like And EGF Like Domains 2; Venous Malformations Multiple Cutaneous And Mucosal

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
1.56-100ng/mL
Sensitivity:
0.52ng/mL
Specificity:
This kit recognizes Human Tie2 in samples. No significant cross-reactivity or interference between Human Tie2 and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Immunology
  • Human TEK Tyrosine Kinase, Endothelial (Tie2) ELISA Kit
  • Human TEK Tyrosine Kinase, Endothelial (Tie2) ELISA Kit
  • Human TEK Tyrosine Kinase, Endothelial (Tie2) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| Tie2

Angiopoietin-1 receptor also known as CD22B (cluster of differentiation 22B) is a protein encoded by the TEK gene. Also known as TIE2, it is an angiopoietin receptor. This receptor possesses a unique extracellular domain containing 2 immunoglobulin-like loops separated by 3 epidermal growth factor-like repeats that are connected to 3 fibronectin type III-like repeats. The ligand for the receptor is angiopoietin-1. TEK has also been suggested as a marker for nucleus pulposus progenitor cells, from the intervertebral disc, which upon activation by Angiopoietin-1 starts to multiply and differentiate.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 100 50 25 12.5 6.25 3.12 1.56 0
OD 2.1 1.46 0.95 0.61 0.4 0.28 0.22 0.08
Corrected OD 2.02 1.38 0.87 0.53 0.32 0.2 0.14 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human Tie2 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-102 96
Plasma(n=8) 92-105 102
Cell Culture Supernates(n=8) 96-108 104





 

| Linearity

Samples were spiked with high concentrations of Human Tie2 and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Culture Supernates(n=4)
1 : 2 Range (%) 84-95 88-96 90-110
Average (%) 91 93 96
1 : 4 Range (%) 89-103 88-108 105-115
Average (%) 94 98 108







 