Product Name:

Human Mannose Receptor C Type 1 (MRC1) ELISA Kit

Cat.#:

JL46837

Regulatory Status:
Alternative:

CLEC13D, CLEC13DL, MRC1L1; C-type lectin domain family 13 member D-like; Macrophage mannose receptor 1-like protein 1

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.312-20ng/mL
Sensitivity:
0.16ng/mL
Specificity:
This kit recognizes human MRC1 in samples. No significant cross-reactivity or interference between human MRC1 and analogues was observed.
Sample Type:
serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Immunology
  • Human Mannose Receptor C Type 1 (MRC1) ELISA Kit
  • Human Mannose Receptor C Type 1 (MRC1) ELISA Kit
  • Human Mannose Receptor C Type 1 (MRC1) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| MRC1

The mannose receptor (Cluster of Differentiation 206, CD206) is a C-type lectin primarily present on the surface of macrophages, immature dendritic cells and liver sinusoidal endothelial cells, but is also expressed on the surface of skin cells such as human dermal fibroblasts and keratinocytes.[1][2] It is the first member of a family of endocytic receptors that includes Endo180 (CD280), M-type PLA2R, and DEC-205 (CD205).
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 20 10 5 2.5 1.25 0.625 0.312 0
OD 2.04 1.4 0.89 0.64 0.42 0.27 0.21 0.06
Corrected OD 1.98 1.34 0.83 0.58 0.36 0.21 0.15 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of human MRC1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 85-101 97
Plasma(n=8) 92-105 102
Cell Culture Supernates(n=8) 96-108 105





 

| Linearity

Samples were spiked with high concentrations of human MRC1 and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Culture Supernates(n=4)
1 : 2 Range (%) 84-95 88-96 89-110
Average (%) 91 93 96
1 : 4 Range (%) 88-103 87-108 105-114
Average (%) 93 98 108







 