400-0066-400

Human Glutathione Peroxidase 1 (GPX1) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human Glutathione Peroxidase 1 (GPX1) ELISA Kit
Product Name:

Human Glutathione Peroxidase 1 (GPX1) ELISA Kit

Cat.#:

JL26307

Regulatory Status:
Alternative:

GSHPX1; Cellular glutathione peroxidase

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
39.06-2500pg/mL
Sensitivity:
14.75pg/mL
Specificity:
This kit recognizes Human GPX1 in samples. No significant cross-reactivity or interference between Human GPX1 and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Cancer
  • Human Glutathione Peroxidase 1 (GPX1) ELISA Kit
  • Human Glutathione Peroxidase 1 (GPX1) ELISA Kit
  • Human Glutathione Peroxidase 1 (GPX1) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| GPX1

Glutathione peroxidase 1, also known as GPx1, is an enzyme encoded by the GPX1 gene. This gene encodes a member of the glutathione peroxidase family. Glutathione peroxidase functions in the detoxification of hydrogen peroxide, and is one of the most important antioxidant enzymes in humans. The protein encoded by this gene forms a homotetramer structure. GPX1 is ubiquitously expressed in many tissues, where it protects cells from oxidative stress.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (pg/mL) 2500 1250 625 312.5 156.25 78.12 39.06 0
OD 2.26 1.45 0.85 0.55 0.36 0.27 0.19 0.07
Corrected OD 2.19 1.38 0.78 0.48 0.29 0.2 0.12 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human GPX1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-101 96
Plasma(n=8) 92-105 102
Cell Culture Supernates(n=8) 96-108 105





 

| Linearity

Samples were spiked with high concentrations of Human GPX1 and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Culture Supernates(n=4)
1 : 2 Range (%) 84-95 88-96 90-110
Average (%) 91 93 96
1 : 4 Range (%) 89-103 87-108 105-115
Average (%) 94 98 109







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
360 / 360
96T
1
460 / 460
Top