400-0066-400

Human Luteinizing Hormone (LH) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human Luteinizing Hormone (LH) ELISA Kit
Product Name:

Human Luteinizing Hormone (LH) ELISA Kit

Cat.#:

JL12917

Regulatory Status:
Alternative:

ICSH; Lutropin; Interstitial Cell Stimulating Hormone

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.156-10mIU/mL
Sensitivity:
0.074mIU/mL
Specificity:
This kit recognizes Human LH in samples. No significant cross-reactivity or interference between Human LH and analogues was observed.
Sample Type:
serum, plasma and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Neuroscience
  • Human Luteinizing Hormone (LH) ELISA Kit
  • Human Luteinizing Hormone (LH) ELISA Kit
  • Human Luteinizing Hormone (LH) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| LH

Luteinizing hormone (LH) is a hormone produced by gonadotropic cells in the anterior pituitary gland. The production of LH is regulated by gonadotropin-releasing hormone (GnRH) from the hypothalamus. In females, an acute rise of LH known as an LH surge, triggers ovulation and development of the corpus luteum. In males, where LH had also been called interstitial cell–stimulating hormone (ICSH), it stimulates Leydig cell production of testosterone. It acts synergistically with follicle-stimulating hormone (FSH).
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (mIU/mL) 10 5 2.5 1.25 0.625 0.312 0.156 0
OD 2.45 1.88 1.46 1.13 0.76 0.49 0.38 0.08
Corrected OD 2.37 1.8 1.38 1.05 0.68 0.41 0.3 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human LH spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 85-101 96
Plasma(n=8) 92-105 102





 

| Linearity

Samples were spiked with high concentrations of Human LH and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4)
1 : 2 Range (%) 84-95 88-96
Average (%) 91 93
1 : 4 Range (%) 89-103 87-108
Average (%) 94 98







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
360 / 360
96T
1
460 / 460
Top