400-0066-400

Human Anti-Type I Interferon Antibodies (IFN-α; IFN-ω-Ab) ELISA Kit

Home  >  Indirect ELSA Kits  >  Human Anti-Type I Interferon Antibodies (IFN-α; IFN-ω-Ab) ELISA Kit
Product Name:

Human Anti-Type I Interferon Antibodies (IFN-α; IFN-ω-Ab) ELISA Kit

Cat.#:

JL55414

Regulatory Status:
Alternative:

anti-IFN-α; IFN-ω-Ab

Detection Method:
ELISA
Assay Type:
Indirect
Detection Range:
0.31-20ng/mL
Sensitivity:
0.14ng/mL
Specificity:
This kit recognizes Human anti-IFN-α;IFN-ω-Ab in samples. No significant cross-reactivity or interference between Human anti-IFN-α;IFN-ω-Ab and analogues was observed.
Sample Type:
Serum, plasma
Assay Procedure:
Assay Procedure
Research Area:
Immunology
  • Human Anti-Type I Interferon Antibodies (IFN-α; IFN-ω-Ab) ELISA Kit
  • Human Anti-Type I Interferon Antibodies (IFN-α; IFN-ω-Ab) ELISA Kit
  • Human Anti-Type I Interferon Antibodies (IFN-α; IFN-ω-Ab) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| IFN-α;IFN-ω-Ab

Human anti-type I interferon antibodies (anti-IFN-α/ω-Ab) are a class of autoantibodies directed against type I interferons, predominantly the IFN-α and IFN-ω subtypes. Their most prominent clinical value lies in serving as highly specific biomarkers for autoimmune polyendocrine syndrome type 1 (APS-1): high-titer anti-IFN-ω antibodies are detectable in over 90% of APS-1 patients, and the positivity rate for anti-IFN-α antibodies approaches nearly 100%, whereas these antibodies are virtually absent in healthy individuals.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 20 10 5 2.5 1.25 0.62 0.31 0
OD 2.38 1.28 0.72 0.38 0.21 0.13 0.08 0.03
Corrected OD 2.35 1.25 0.69 0.35 0.18 0.1 0.05 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human IFN-α;IFN-ω-Ab spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 87-114 101
Plasma(n=8) 85-111 98





 

| Linearity

Samples were spiked with high concentrations of Human IFN-α;IFN-ω-Ab and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4)
1 : 2 Range (%) 92-109 84-110
Average (%) 97 95
1 : 4 Range (%) 91-106 88-107
Average (%) 97 101







 