Product Name:
Human Folate Receptor Autoantibody (FRAAs) ELISA Kit
Cat.#:
JL18368
Regulatory Status:
Alternative:
FRAAs; FRα-Ab; Folate Receptor Autoantibodies
Detection Method:
ELISA
Assay Type:
Indirect
Detection Range:
1.56-100ng/mL
Sensitivity:
0.77ng/mL
Specificity:
This kit recognizes Human FRAAs in samples. No significant cross-reactivity or interference between Human FRAAs and analogues was observed.
Sample Type:
Serum, plasma and other biological fluids
Assay Procedure:
Research Area:
Immunology
Product Datasheets
| FRAAs
The folate receptor is a glycosylphosphatidylinositol-coupling protein. Except for certain tissues, folate receptor expression levels are low in normal tissues but overexpressed on the surface of many tumor cells. The folate receptor exhibits high affinity for folate and its derivatives. Leveraging this property, imaging agents, therapeutic drugs, and other molecules can be conjugated to folate for targeted delivery to tumor cells. This enables applications in tumor imaging diagnostics—such as nuclear medicine imaging, magnetic resonance imaging (MRI), and fluorescence imaging—as well as tumor therapies including chemotherapy, isotope therapy, immunotherapy, antisense nucleic acid therapy, and gene therapy.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (ng/mL) | 100 | 50 | 25 | 12.5 | 6.25 | 3.12 | 1.56 | 0 |
| OD | 2.66 | 1.48 | 0.77 | 0.4 | 0.25 | 0.15 | 0.11 | 0.05 |
| Corrected OD | 2.61 | 1.43 | 0.72 | 0.35 | 0.2 | 0.1 | 0.06 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human FRAAs spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 87-104 | 98 |
| Plasma(n=8) | 99-106 | 101 |
| Linearity
Samples were spiked with high concentrations of Human FRAAs and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | ||
| 1 : 2 | Range (%) | 82-95 | 89-102 |
| Average (%) | 92 | 95 | |
| 1 : 4 | Range (%) | 89-104 | 87-110 |
| Average (%) | 97 | 99 |
Size
Quantity
Price ($)
48T
1
460 / 460
96T
1
760 / 760