Product Name:
Mouse Beta-Endorphin (β-EP) ELISA Kit
Cat.#:
JL10523
Regulatory Status:
Alternative:
b-EP
Detection Method:
ELISA
Assay Type:
Competitive
Detection Range:
15.6-1000pg/mL
Sensitivity:
7.6pg/mL
Specificity:
This kit recognizes Mouse β-EP in samples. No significant cross-reactivity or interference between Mouse β-EP and analogues was observed.
Sample Type:
Serum;plasma;cerebrospinal fluid and other biological fluids
Assay Procedure:
Research Area:
Signal Transduction
Product Datasheets
| β-EP
β-Endorphin (beta-endorphin) is an endogenous opioid neuropeptide and peptide hormone that is produced in certain neurons within the central nervous system and peripheral nervous system.It is one of three endorphins that are produced in humans, the others being α-endorphin and γ-endorphin.Function of β-endorphin has been known to be associated with hunger, thrill, pain, maternal care, sexual behavior, and reward cognition. In the broadest sense, β-endorphin is primarily utilized in the body to reduce stress and maintain homeostasis. In behavioral research, studies have shown that β-endorphin is released via volume transmission into the ventricular system in response to a variety of stimuli, and novel stimuli in particular.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (pg/mL) | 1000 | 500 | 250 | 125 | 62.5 | 31.2 | 15.6 | 0 |
| OD | 0.216 | 0.298 | 0.417 | 0.669 | 0.955 | 1.405 | 1.79 | 1.865 |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Mouse β-EP spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 81-101 | 96 |
| Plasma(n=8) | 92-105 | 101 |
| Cerebrospinal Fluid(n=8) | 97-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Mouse β-EP and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | Cerebrospinal Fluid(n=4) | ||
| 1 : 2 | Range (%) | 83-95 | 88-96 | 90-110 |
| Average (%) | 91 | 93 | 96 | |
| 1 : 4 | Range (%) | 89-103 | 87-108 | 105-115 |
| Average (%) | 93 | 98 | 110 |
Size
Quantity
Price ($)
48T
1
360 / 360
96T
1
460 / 460