Detection Range:
31.25-2000ng/mL
Sensitivity:
15.3ng/mL
Specificity:
This kit recognizes Human IgG3 in samples. No significant cross-reactivity or interference between Human IgG3 and analogues was observed.
Sample Type:
Serum, plasma and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Immunology
  • Human Immunoglobulin G3 (IgG3) ELISA Kit
  • Human Immunoglobulin G3 (IgG3) ELISA Kit
  • Human Immunoglobulin G3 (IgG3) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| IgG3

Immunoglobulin G3 (IgG3) is nephritogenic and avidly fixes complement; its high molecular weight and anionic charge favor its localization to the glomerular capillary wall. The main feature of IgG3 is the long hinge connecting the Fab domains and the Fc portion of the molecule. The hinge of IgG3 is 62 amino acids long, more than four times that of IgG1, and contains 11 disulfide bonds. In addition, depending on the allotype of IgG3, its half-life is substantially lower than that of the other subclasses.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 2000 1000 500 250 125 62.5 31.25 0
OD 0.188 0.27 0.389 0.642 0.929 1.38 1.766 1.842
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human IgG3 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 88-102 98
Plasma(n=8) 94-107 104





 

| Linearity

Samples were spiked with high concentrations of Human IgG3 and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4)
1 : 2 Range (%) 82-101 88-109
Average (%) 95 99
1 : 4 Range (%) 89-103 87-111
Average (%) 93 98







 