Human 6-hydroxymelatonin Sulfate (6-SMT) ELISA Kit

Home  >  Competitive ELISA Kits  >  Human 6-hydroxymelatonin Sulfate (6-SMT) ELISA Kit
Product Name:

Human 6-hydroxymelatonin Sulfate (6-SMT) ELISA Kit

Cat.#:

JL13518

Regulatory Status:
Alternative:

6-hydroxymelatonin sulfate

Detection Method:
ELISA
Assay Type:
Competitive
Detection Range:
3.9-250pg/mL
Sensitivity:
1.9pg/mL
Specificity:
This kit recognizes Human 6-SMT in samples. No significant cross-reactivity or interference between Human 6-SMT and analogues was observed.
Sample Type:
Serum;plasma and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Immunology
  • Human 6-hydroxymelatonin Sulfate (6-SMT) ELISA Kit
  • Human 6-hydroxymelatonin Sulfate (6-SMT) ELISA Kit
  • Human 6-hydroxymelatonin Sulfate (6-SMT) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| 6-SMT

Melatonin, an indoleamine, is a natural compound produced by various organisms, including bacteria and eukaryotes.Melatonin is metabolized with an elimination half-life ranging from 20 to 50 minutes.The primary metabolic pathway transforms melatonin into 6-hydroxymelatonin, which is then conjugated with sulfate and excreted in urine as a waste product.6-hydroxymelatonin Sulfate (6-SMT) is the main excretion form of melatonin in urine.6-SMT as a stable terminal metabolite for the assessment of long-term melatonin secretion levels (e.g., in sleep disorders, depression diagnosis).
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (pg/mL) 250 125 62.5 31.2 15.6 7.8 3.9 0
OD 0.19 0.24 0.31 0.55 0.97 1.38 1.96 2.52
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human 6-SMT spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 81-101 96
Plasma(n=8) 92-105 101
Cell Culture Supernates(n=8) 97-107 104





 

| Linearity

Samples were spiked with high concentrations of Human 6-SMT and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Culture Supernates(n=4)
1 : 2 Range (%) 83-95 88-96 90-110
Average (%) 91 93 96
1 : 4 Range (%) 89-105 87-108 105-115
Average (%) 94 98 110







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
360 / 360
96T
1
460 / 460