Product Name:
Human Thymic Factor (TF) ELISA Kit
Cat.#:
JL52524
Regulatory Status:
Alternative:
FTS; Facteur Thymique Serique; Thymulin
Detection Method:
ELISA
Assay Type:
Competitive
Detection Range:
46.8-3000pg/mL
Sensitivity:
23.6pg/mL
Specificity:
This kit recognizes Human TF in samples. No significant cross-reactivity or interference between Human TF and analogues was observed.
Sample Type:
Serum;plasma and other biological fluids
Assay Procedure:
Research Area:
Cardiovascular Biology
Product Datasheets
| TF
Thymulin (also known as thymic factor or facteur thymique serique) is a nonapeptide produced by two distinct epithelial populations in the thymus.It requires zinc for biological activity. Its peptide sequence is H-Pyr-Ala-Lys-Ser-Gln-Gly-Gly-Ser-Asn-OH.The hormone is believed to be involved in T-cell differentiation and enhancement of T and NK cell actions.Besides this rather paracrine or auto-organic effects on the thymus dependent immune system, thymulin seems to have neuroendocrine effects as well. There exist bidirectional interactions between thymic epithelium and the hypothalamus-pituitary axis (for example, thymulin follows a circadian rhythm and physiologically elevated ACTH levels correlate positively with thymulin plasma levels and vice versa).| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (pg/mL) | 3000 | 1500 | 750 | 375 | 187.5 | 93.7 | 46.8 | 0 |
| OD | 0.188 | 0.27 | 0.389 | 0.642 | 0.929 | 1.38 | 1.766 | 1.842 |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human TF spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 81-101 | 96 |
| Plasma(n=8) | 92-105 | 101 |
| Linearity
Samples were spiked with high concentrations of Human TF and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | ||
| 1 : 2 | Range (%) | 83-95 | 88-96 |
| Average (%) | 91 | 93 | |
| 1 : 4 | Range (%) | 89-104 | 87-108 |
| Average (%) | 93 | 98 |
Size
Quantity
Price ($)
48T
1
360 / 360
96T
1
460 / 460