Product Name:
Human Platelet Activating Factor (PAF) ELISA Kit
Cat.#:
JL10406
Regulatory Status:
Alternative:
PAF-Acether; AGEPC; Acetyl-Glyceryl-Ether-Phosphorylcholine
Detection Method:
ELISA
Assay Type:
Competitive
Detection Range:
46.87-3000pg/mL
Sensitivity:
23pg/mL
Specificity:
This kit recognizes Human PAF in samples. No significant cross-reactivity or interference between Human PAF and analogues was observed.
Sample Type:
Serum;plasma and other biological fluids
Assay Procedure:
Research Area:
Epigenetics
Product Datasheets
| PAF
Platelet-activating factor(PAF) is a potent phospholipid activator and mediator of many leukocyte functions, platelet aggregation and degranulation, inflammation, and anaphylaxis. It is also involved in changes to vascular permeability, the oxidative burst, chemotaxis of leukocytes, as well as augmentation of arachidonic acid metabolism in phagocytes.PAF is produced by a variety of cells, but especially those involved in host defense, such as platelets, endothelial cells, neutrophils, monocytes, and macrophages. PAF is continuously produced by these cells but in low quantities and production is controlled by the activity of PAF acetylhydrolases. It is produced in larger quantities by inflammatory cells in response to specific stimuli.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (pg/mL) | 3000 | 1500 | 750 | 375 | 187.5 | 93.75 | 46.87 | 0 |
| OD | 0.19 | 0.24 | 0.31 | 0.55 | 0.97 | 1.38 | 1.96 | 2.52 |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human PAF spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 81-101 | 96 |
| Plasma(n=8) | 92-105 | 101 |
| Linearity
Samples were spiked with high concentrations of Human PAF and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | ||
| 1 : 2 | Range (%) | 83-95 | 88-96 |
| Average (%) | 91 | 93 | |
| 1 : 4 | Range (%) | 89-104 | 87-108 |
| Average (%) | 93 | 98 |
Size
Quantity
Price ($)
48T
1
360 / 360
96T
1
460 / 460