400-0066-400

Sulfuric Acid Dehydroepiandrosterone (DHEAS) ELISA Kit

Home  >  Competitive ELISA Kits  >  Sulfuric Acid Dehydroepiandrosterone (DHEAS) ELISA Kit
Product Name:

Sulfuric Acid Dehydroepiandrosterone (DHEAS) ELISA Kit

Cat.#:

JL58278

Regulatory Status:
Alternative:

DHEAS

Detection Method:
ELISA
Assay Type:
Competitive
Detection Range:
0.15-10ng/mL
Sensitivity:
0.07ng/mL
Specificity:
This kit recognizes DHEAS in samples. No significant cross-reactivity or interference between DHEAS and analogues was observed.
Sample Type:
Serum;plasma and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Metabolism
  • Sulfuric Acid Dehydroepiandrosterone (DHEAS) ELISA Kit
  • Sulfuric Acid Dehydroepiandrosterone (DHEAS) ELISA Kit
  • Sulfuric Acid Dehydroepiandrosterone (DHEAS) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| DHEAS

Dehydroepiandrosterone sulfate, abbreviated as DHEA sulfate or DHEA-S, also known as androstenolone sulfate, is an endogenous androstane steroid that is produced by the adrenal cortex.It is the 3β-sulfate ester and a metabolite of dehydroepiandrosterone (DHEA) and circulates in far greater relative concentrations than DHEA.The steroid is hormonally inert and is instead an important neurosteroid and neurotrophin.DHEA-S can be converted back into DHEA via steroid sulfatase (STS).Similarly to other conjugated steroids, DHEA-S is devoid of hormonal activity, lacking affinity for the steroid hormone receptors.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 10 5 2.5 1.25 0.62 0.31 0.15 0
OD 0.19 0.24 0.31 0.55 0.97 1.38 1.96 2.52
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of DHEAS spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 81-101 96
Plasma(n=8) 92-105 102





 

| Linearity

Samples were spiked with high concentrations of DHEAS and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4)
1 : 2 Range (%) 83-95 88-96
Average (%) 91 93
1 : 4 Range (%) 89-104 87-108
Average (%) 93 98







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
420 / 420
96T
1
640 / 640