400-0066-400

Epinephrine (EPI) ELISA Kit

Home  >  Competitive ELISA Kits  >  Epinephrine (EPI) ELISA Kit
Product Name:

Epinephrine (EPI) ELISA Kit

Cat.#:

JL53762

Regulatory Status:
Alternative:

Epinephrine;adrenaline;3,4,β-Trihydroxy-N-methylphenethylamine

Detection Method:
ELISA
Assay Type:
Competitive
Detection Range:
31.2-2000pg/mL
Sensitivity:
15.8pg/mL
Specificity:
This kit recognizes EPI in samples. No significant cross-reactivity or interference between EPI and analogues was observed.
Sample Type:
Serum;plasma;tissue homogenates;cell lysates;cell culture supernates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Neuroscience
  • Epinephrine (EPI) ELISA Kit
  • Epinephrine (EPI) ELISA Kit
  • Epinephrine (EPI) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| EPI

Adrenaline, also known as epinephrine, is a hormone and medication which is involved in regulating visceral functions (e.g., respiration).It appears as a white microcrystalline granule.Adrenaline is normally produced by the adrenal glands and by a small number of neurons in the medulla oblongata.It plays an essential role in the fight-or-flight response by increasing blood flow to muscles, heart output by acting on the SA node,pupil dilation response, and blood sugar level.It does this by binding to alpha and beta receptors.It is found in many animals, including humans, and some single-celled organisms.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (pg/mL) 2000 1000 500 250 125 62.5 31.2 0
OD 0.216 0.298 0.417 0.669 0.955 1.405 1.79 1.865
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of EPI spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 81-101 96
Plasma(n=8) 92-105 101
Cell Culture Supernates(n=8) 97-108 104





 

| Linearity

Samples were spiked with high concentrations of EPI and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Culture Supernates(n=4)
1 : 2 Range (%) 83-95 88-96 90-110
Average (%) 91 93 96
1 : 4 Range (%) 89-104 87-108 105-115
Average (%) 97 97 110







 