Product Name:
Melatonin (MT) ELISA Kit
Cat.#:
JL54424
Regulatory Status:
Alternative:
N-Acetyl-5-Methoxytryptamine
Detection Method:
ELISA
Assay Type:
Competitive
Detection Range:
7.8-500pg/mL
Sensitivity:
3.4pg/mL
Specificity:
This kit recognizes MT in samples. No significant cross-reactivity or interference between MT and analogues was observed.
Sample Type:
Serum;plasma and other biological fluids
Assay Procedure:
Research Area:
Metabolism
Product Datasheets
| MT
Melatonin, an indoleamine, is a natural compound produced by various organisms, including bacteria and eukaryotes.This compound was later identified as a hormone secreted in the brain during the night, playing a crucial role in regulating the sleep-wake cycle, also known as the circadian rhythm, in vertebrates.In vertebrates, melatonin's functions extend to synchronizing sleep-wake cycles, encompassing sleep-wake timing and blood pressure regulation, as well as controlling seasonal rhythmicity (circannual cycle), which includes reproduction, fattening, molting, and hibernation.Its effects are mediated through the activation of melatonin receptors and its role as an antioxidant.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (pg/mL) | 500 | 250 | 125 | 62.5 | 31.2 | 15.6 | 7.8 | 0 |
| OD | 0.216 | 0.298 | 0.417 | 0.669 | 0.955 | 1.405 | 1.79 | 1.865 |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of MT spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 81-101 | 96 |
| Plasma(n=8) | 92-105 | 102 |
| Linearity
Samples were spiked with high concentrations of MT and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | ||
| 1 : 2 | Range (%) | 83-95 | 88-96 |
| Average (%) | 91 | 93 | |
| 1 : 4 | Range (%) | 89-104 | 87-108 |
| Average (%) | 93 | 98 |
Size
Quantity
Price ($)
48T
1
396 / 396
96T
1
516 / 516