Product Name:
Streptomycin (Streptomycin) ELISA Kit
Cat.#:
JL55000
Regulatory Status:
Alternative:
Streptomycin
Detection Method:
ELISA
Assay Type:
Competitive
Detection Range:
0.156-10ng/mL
Sensitivity:
0.059ng/mL
Specificity:
This kit recognizes Streptomycin in samples. No significant cross-reactivity or interference between Streptomycin and analogues was observed.
Sample Type:
Serum;plasma;tissue homogenates;cell lysates;cell culture supernates and other biological fluids
Assay Procedure:
Research Area:
Pesticide Residues & Drug Metabolism
Product Datasheets
| Streptomycin
Streptomycin is an antibiotic medication used to treat a number of bacterial infections,including tuberculosis, Mycobacterium avium complex, endocarditis, brucellosis, Burkholderia infection, plague, tularemia, and rat bite fever.For active tuberculosis it is often given together with isoniazid, rifampicin, and pyrazinamide.It is administered by injection into a vein or muscle.Streptomycin is an aminoglycoside.It works by blocking the ability of 30S ribosomal subunits to make proteins, which results in bacterial death.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (ng/mL) | 10 | 5 | 2.5 | 1.25 | 0.625 | 0.312 | 0.156 | 0 |
| OD | 0.188 | 0.27 | 0.389 | 0.642 | 0.929 | 1.38 | 1.766 | 1.842 |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Streptomycin spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 85-101 | 97 |
| Plasma(n=8) | 91-105 | 102 |
| Cell Culture Supernates(n=8) | 96-108 | 104 |
| Linearity
Samples were spiked with high concentrations of Streptomycin and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | Cell Culture Supernates(n=4) | ||
| 1 : 2 | Range (%) | 85-95 | 88-96 | 90-111 |
| Average (%) | 91 | 93 | 96 | |
| 1 : 4 | Range (%) | 89-104 | 87-108 | 105-116 |
| Average (%) | 94 | 98 | 109 |
Size
Quantity
Price ($)
48T
1
500 / 500
96T
1
640 / 640