Product Name:
Bovine Estriol (E3) ELISA Kit
Cat.#:
JL16484
Regulatory Status:
Alternative:
oestriol;Estra-1,3,5(10)-triene-3,16α,17β-triol;Oestriol; E3; Estratriol;Theelol;Trihydroxyestrin; Trihydroxyoestrin;16α-Hydroxyestradiol
Detection Method:
ELISA
Assay Type:
Competitive
Detection Range:
1.56-100ng/mL
Sensitivity:
0.72ng/mL
Specificity:
This kit recognizes Bovine E3 in samples. No significant cross-reactivity or interference between Bovine E3 and analogues was observed.
Sample Type:
Serum;plasma;tissue homogenates;cell lysates;cell culture supernates and other biological fluids
Assay Procedure:
Research Area:
Signal Transduction
Product Datasheets
| E3
Estriol (E3), also spelled oestriol, is a steroid, a weak estrogen, and a minor female sex hormone.It is one of three major endogenous estrogens, the others being estradiol and estrone.Levels of estriol in women who are not pregnant are almost undetectable.However, during pregnancy, estriol is synthesized in very high quantities by the placenta and is the most produced estrogen in the body by far,although circulating levels of estriol are similar to those of other estrogens due to a relatively high rate of metabolism and excretion.Relative to estradiol, both estriol and estrone have far weaker activity as estrogens.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (ng/mL) | 100 | 50 | 25 | 12.5 | 6.25 | 3.12 | 1.56 | 0 |
| OD | 0.188 | 0.27 | 0.389 | 0.642 | 0.929 | 1.38 | 1.766 | 1.842 |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Bovine E3 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 81-101 | 96 |
| Plasma(n=8) | 92-105 | 101 |
| Cell Culture Supernates(n=8) | 97-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Bovine E3 and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | Cell Culture Supernates(n=4) | ||
| 1 : 2 | Range (%) | 83-95 | 88-96 | 90-110 |
| Average (%) | 91 | 93 | 96 | |
| 1 : 4 | Range (%) | 89-104 | 87-108 | 105-115 |
| Average (%) | 93 | 98 | 110 |
Size
Quantity
Price ($)
48T
1
396 / 396
96T
1
576 / 576